Page C.2
<y_divisions> = <unsigned_integer>
Number of y divisions in which the waveform
is displayed on the instrument screen.
<x_divisions> = <unsigned_integer>
Number of x divisions in which the waveform
is displayed on the instrument screen.
<y_scale> = <float>
Number of units per y division.
<x_scale> = <float>
Number of units per x division.
<y_step> = <binary_character>
Specifies in which scale the <y_scale> is set
by the instrument:
1 = 1-2-5 range
2 = 1-2-4 range
<x_step> = <binary_character>
Specifies in which scale the <x_scale> is set
by the instrument:
1 = 1-2-5 range
3 = record range
4 = variable range
<y_zero> = <float>
Measurement value for the samples with value
zero (0) that you can see as offset value.
<x_zero> = <float>
This field specifies the x-offset of the
first sample in <trace_samples>. (is time
between trigger moment and first sample.)
<y_resolution> = <float>
This field contains the value that
represents the step between two
consecutive sample values or in other
words the step per least significant bit.
<x_resolution> = <float>
This field contains the value (seconds)
that represents the distance between two
samples. (is time between two samples.)
In the case of an FFT-trace, this value
is the frequency of the fundamental (Hz).
Summary of Contents for ScopeMeter 190 Series
Page 40: ...Page 3 29 Response Syntax acknowledge cr See an example for this command under QUERY PRINT QP...
Page 42: ...to the ScopeMeter...
Page 64: ......
Page 68: ...Resp Count Resp Count 2 PRINT x_divisions X Divisions...
Page 70: ...PRINT x_step X Step...
Page 72: ...nb nb ASC MID Resp Resp Count 1 1 ELSE...
Page 74: ......
Page 76: ...PRINT SumCheck so far MOD 256 SumCheck1 MOD 256 CLOSE END...
Page 78: ...END IF...
Page 80: ...END IF NEXT j...
Page 83: ...Resp Count Resp Count Sample Bytes...