Page 3.25
=============================================================
INSTRUMENT STATUS IS
-------------------------------------------------------------
Purpose:
Queries the contents of the ScopeMeter’s status register.
The returned value reflects the present operational status
of the ScopeMeter. This is a 16-bit word, presented as an
integer value, where each bit represents the Boolean value
of a related event.
Command Syntax:
IS<cr>
Response Syntax:
<acknowledge><cr>[<status><cr>]
where,
<status> = integer value 0 to 65535
<status>
Bit Value Status Description
----------------------------------------------------------
0 1 Maintenance mode
1 2 Charging
2 4 Recording
3 8 AutoRanging
4 16 Remote
5 32 Battery Connected
6 64 Power (Net) Adapter connected
7 128 Calibration necessary
8 256 Instrument Held (HOLD status)
9 512 Pre Calibration busy
10 1024 Pre Calibration valid
11 2048 Replay buffer full
12 4096 Triggered
13 8192 Instrument On
14 16384 Instrument Reset occurred
15 32768 Next <status> value available