Page C.1
=============================================================
APPENDIX C WAVEFORM DATA
-------------------------------------------------------------
The waveform data that is received from the QW (Query
Waveform) query, consists of the following data.
<trace_admin>,<trace_samples>
where,
<trace_admin> = #0<block_header><block_length><trace_result>
<y_unit><x_unit><y_divisions><x_divisions>
<y_scale><x_scale><y-step><x_step><y_zero>
<x_zero><y_resolution><x_resolution><y_at_0>
<x_at_0><date_stamp><time_stamp><check_sum>
where,
<block_header> = <binary_character>
Possible values: 144 and 0.
The value 0 is returned when also the
<trace_samples> data block is requested.
<block_length> = <unsigned_integer>
This value gives the number of bytes that
are transmitted after the <block_length>
and before the <check_sum>.
<trace_result> = <binary_character>
If bit 0 is set (decimal value 1) the trace is a direct
result of a trace acquisition.
If bit 1 is set (decimal value 2) the trace is a result
of the TrendPlot function (recording numerical results).
If bit 2 is set (decimal value 4) either the trace itself
is an envelope trace, or an envelope trace is available.
If bit 3 is set (decimal value 8) either the trace itself
is a reference trace, or a reference trace is available.
If bit 4 is set (decimal value 16) either the trace itself
is a mathematics trace, or a mathematics trace is
available.
Note: This <trace_result> information is not available in
all instrument types/versions.
<y_unit> = <unit>
<x_unit> = <unit>
The <unit> is a <binary_character> which
value represents the unit:
None = 0
<Volt> = 1
<Ampere> = 2
<Ohm> = 3
<Watt> = 4
<Farad> = 5
<Kelvin> = 6
<seconds> = 7
Summary of Contents for ScopeMeter 190 Series
Page 40: ...Page 3 29 Response Syntax acknowledge cr See an example for this command under QUERY PRINT QP...
Page 42: ...to the ScopeMeter...
Page 64: ......
Page 68: ...Resp Count Resp Count 2 PRINT x_divisions X Divisions...
Page 70: ...PRINT x_step X Step...
Page 72: ...nb nb ASC MID Resp Resp Count 1 1 ELSE...
Page 74: ......
Page 76: ...PRINT SumCheck so far MOD 256 SumCheck1 MOD 256 CLOSE END...
Page 78: ...END IF...
Page 80: ...END IF NEXT j...
Page 83: ...Resp Count Resp Count Sample Bytes...