Page 3.35
Response Syntax 1:
<acknowledge><cr>[<reading>{,<reading>}<cr>]
where,
<reading> = <no>,<valid>,<source>,<unit>,<type>,<pres>,<resol>
<no> see Command Syntax 2
<valid> validity of the reading:
1 reading valid
0 reading non-valid
<source> source of the reading:
1 Voltage channel: Input A (Scope mode)
2 Ampere channel: Input B (Scope mode)
3 Input external: COM & V/Ohm/Diode (Meter mode)
12 Input_AB (Phase A over B): A>B (Scope mode),
or M (Mathematics A+B, A-B or AxB)
21 Input_BA (Phase B over A): B>A (Scope mode)
<unit> unit of the reading:
0 None (off)
1 Volt
2 Ampere
3 Ohm
4 Watt
5 Farad
6 Kelvin
7 seconds
8 hours
9 days
10 Hertz
11 Degrees
12 Celsius
13 Fahrenheit
14 percentage (%)
15 dBm 50 Ohm
16 dBm 600 Ohm
17 dBVolt
18 dBAmpere
19 dBWatt
20 Volt * Ampere Reactive (VAR)
21 Volt * Ampere (VA)
<type> reading characteristic of the measurement:
0 None
1 Mean
2 Rms
3 True rms
4 Peak peak
5 Peak maximum
6 Peak minimum
7 Crest factor
8 Period
9 Duty cycle negative
10 Duty cycle positive
11 Frequency
12 Pulse width negative
Summary of Contents for ScopeMeter 190 Series
Page 40: ...Page 3 29 Response Syntax acknowledge cr See an example for this command under QUERY PRINT QP...
Page 42: ...to the ScopeMeter...
Page 64: ......
Page 68: ...Resp Count Resp Count 2 PRINT x_divisions X Divisions...
Page 70: ...PRINT x_step X Step...
Page 72: ...nb nb ASC MID Resp Resp Count 1 1 ELSE...
Page 74: ......
Page 76: ...PRINT SumCheck so far MOD 256 SumCheck1 MOD 256 CLOSE END...
Page 78: ...END IF...
Page 80: ...END IF NEXT j...
Page 83: ...Resp Count Resp Count Sample Bytes...