7 Connection
7.1 Transducers
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
60
7.1.5
Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced or added, the sensor module must also be replaced or added.
The sensor module is connected to the terminals of the transmitter, see Fig. 7.7.
• Disconnect the transmitter from the power supply.
• Connect the transducer module to the corresponding terminal, see Fig. 7.7. Terminal X2 is assigned to the transducers
of measuring channel A, terminal X3 to the transducers of measuring channel B.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Parameters
.
• Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer must be identical. A wrong or incorrectly connected
sensor module will lead to incorrect measured values or to a measurement failure.
Fig. 7.7:
Sensor module
1 – transducer module (measuring channel A)
2 – transducer module (measuring channel B)
A
B
SNAP
DISP
DISP
MODE
FAST
MUX
NEXT
Q
ON
Q+
Q-
3x Q
OFF
LAN
USB
A+
B-
P1+
P2+
P3+
P4+
P5a
P6a
P7a
S
S
P1-
P2-
P3-
P4-
P5b
P6b
P7b
X2
X3
X_AV
KL1
1
KL12
KL7
KL8
KL9
KL10
KL4
X_AR
X_BV
X_BR
T1A
T1B
S1
T2A
T2B
T3A
T3B
S3
T4B
T4A
T1a
T1b
S1
T2a
T2b
T3a
T3b
S3
T4b
T4a
N(-)
PE
L(+)
AV
AV
S
AGN
ARS
AR
BV
BVS BGN
BRS
BR
1
2
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......