10 Troubleshooting
FLUXUS WD
10.6 Problems with the totalizers
97
UMFLUXUS_WDV1-3EN, 2018-07-16
The entered pipe roughness is not appropriate
The flow velocity of the fluid is outside the measuring range of the transmitter
The measuring point is not appropriate
Check whether a different measuring point provides better results. Because pipes are never rotationally symmetric, the
flow profile is affected.
10.6
Problems with the totalizers
The values of the totalizers are too small
One of the totalizers has reached the upper limit and has to be reset to zero manually.
The sum of the totalizers is not correct
The output sum of both totalizers (throughput
Q
) transmitted via an output is not valid after one of the totalizers has over-
flowed for the first time.
A question mark is displayed after the value of the totalizer
The measurement was temporarily impossible, therefore the totalizer value can be wrong.
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......