13 User modes
FLUXUS WD
13.3 SuperUser Mode and SuperUser ext. mode
111
UMFLUXUS_WDV1-3EN, 2018-07-16
Weighting factor
The weighting factor is used for transducers installed on the same pipe in order to compensate differences between the
measured values of the flow velocity of different channels.
The differences can be caused by profile deformations or transverse flows. These influences can be reduced by averaging
the measured values of several channels. If a channel has a temporarily malfunction, the average will change abruptly. In
order to avoid differences, all channels have to be adjusted with the weighting factor.
The weighting factor for the channel x results from the flow velocity v
x
and the average flow velocity of all channels v
average
:
w
x
=
The weighting factor can be activated in the transmitter.
• Select the menu item
Special settings
in the program branch
Options
.
• Press ENTER until the menu item
Weighting factor
is displayed.
• Select
Off
to measure without the weighting factor.
• Select
Default
if no user-defined inputs are to be made.
• Select
On
to define the value for the weighting factor.
• Press ENTER.
Transducer temperature and transducer temperature violation as diagnostic value
When configuring outputs, the list items
Transducer temp.
and
Transd. temp. violat.
are available in the menu
item
Diagnostic values
. The diagnostic values can either be transmitted via the output of the transmitter or defined as
source of the event trigger. The diagnostic value
Transducer temp.
is available on physical channels only.
• Select
Diagnostic values
as source item.
• Press ENTER.
• Select a list item for the quantity to be output, see Tab. 13.2.
• Press ENTER.
Options\Special settings\Weighting factor
Options\Channel...\Outputs\...\Source item
Tab. 13.2: Source item
Diagnostic values
source item
list item
output
Diagnostic values
Transducer temp.
average temperature of both transducers
Transd. temp. violat.
status information: yes/no
v
average
v
x
--------------------
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......