9 Measurement
9.2 Measurement settings
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
84
• Enter a test value for the selected physical quantity. It has to be within the output range.
• Press ENTER. If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test. Select
Finish
to display the next menu item.
• Press ENTER.
9.2.5.2
Output of a status value
• Select the list item
Options\Outputs\...\Status
.
• Press ENTER.
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If
Other range
is selected, enter the values
Output
MIN
and
Output MAX
.
The output range has to have > 10 % of the max. output
value (
Output MAX
). An error message will be displayed if
the output range is smaller. The next possible value will be
displayed.
Status OK
The status of the output signal is defined which is to be output when measuring a measured value.
• Select in the scroll list the value for
Status OK
.
• Press ENTER.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key
or
further information is displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Options\Channel A\Outputs\...\Enter test value
Source item
Status
Output range
Status OK
Output info
Test signal
Test mea. range
Options\Channel A\Outputs\...\Output range
Options\Channel A\Outputs\...\Status OK
Options\Channel A\Outputs\...\Output info
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......