14 Outputs
FLUXUS WD
14.1 Configuration of a binary output
115
UMFLUXUS_WDV1-3EN, 2018-07-16
If no measurement is carried out, all binary outputs are de-energized, independent of the programmed switching condition.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key
or
further information is displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Output function test
The function of the output can now be tested.
• Connect a multimeter to the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Select
On
to test the energized state of the output.
• Press ENTER.
The value has to be low ohmic.
• Select
Off
to test the de-energized state of the output.
• Press ENTER.
The value has to be high ohmic.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
.
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select in the scroll list
Passive
(idle state) or
Active
.
• Press ENTER.
Depending on the selected switching condition, the value has to be either high or low ohmic.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
Tab. 14.3: Selection of the switching function
property
setting
description
switching function
NO
The binary output is energized if the switching condition is met
(event occurs) and de-energized if the switching condition is not
met (idle state).
NC
The binary output is de-energized if the switching condition is
met (event occurs) and energized if the switching condition is
not met (idle state).
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal B1
Options\Channel A\Outputs\...\Enter test value B1
Options\Channel A\Outputs\...\Test mea. range B1
Options\Outputs\...\Enter test value B1
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......