7 Connection
7.1 Transducers
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
54
7.1.2
Connection of the extension cable to the transmitter
• Remove the blind plug for the connection of the extension cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter, see Fig. Fig. 7.3 and Tab. 7.2.
Tab. 7.1:
Terminal assignment
terminal
connection
AV
transducer
(core)
AVS
transducer (internal shield)
ARS
transducer (inner shield)
AR
transducer
(core)
Fig. 7.2:
Connection of the transducer cable with plastic cable jacket and stripped cable ends to
the transmitter
1 – cap nut
2 – compression part
3 – basic part
4 – external shield brushed back
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
A
B
SNAP
DISP
DISP
MODE
FAST
MUX
NEXT
Q
ON
Q+
Q-
3x Q
OFF
LAN
USB
A+
B-
P1+
P2+
P3+
P4+
P5a
P6a
P7a
S
S
P1-
P2-
P3-
P4-
P5b
P6b
P7b
X2
X3
X_AV
KL1
1
KL12
KL7
KL8
KL9
KL10
KL4
X_AR
X_BV
X_BR
T1A
T1B
S1
T2A
T2B
T3A
T3B
S3
T4B
T4A
T1a
T1b
S1
T2a
T2b
T3a
T3b
S3
T4b
T4a
N(-)
PE
L(+)
AV
AV
S
AGN
ARS
AR
BV
BVS BGN
BRS
BR
70 mm
20 mm
10 mm
4
cable gland
1
2
3
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......