9 Measurement
FLUXUS WD
9.2 Measurement settings
83
UMFLUXUS_WDV1-3EN, 2018-07-16
Measuring range
The sign of the measured value and the measuring range are determined.
• Select
Sign
if the sign of the measured values is to be considered for the output.
• Select
Absolute value
if the sign of the measured values is not to be considered for the output.
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Start of meas. range
is the value assigned to the value
Output MIN
of the output range.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
End of meas. range
is the value assigned to the value
Output MAX
of the output range.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key
or
further information is displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the assignment of the measured value to the output signal. Select
No
to display the next menu item.
• Press ENTER.
Options\Channel A\Outputs\...\Measured values\Absolute value
Options\Channel A\Outputs\...\Start of meas. range
Options\Channel A\Outputs\...\End of meas. range
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......