9 Measurement
9.2 Measurement settings
FLUXUS WD
2018-07-16, UMFLUXUS_WDV1-3EN
86
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing key
or
further information is displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive.
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Yes
to test the output. Select
No
to display the next menu item.
• Press ENTER.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select in the scroll list
Passive
(idle state) or
Active
.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Press key
to return to the program branch
Options
.
9.2.5.4
Switchable current output
• Select the menu item
Outputs
in the program branch
Special functions
.
• Press ENTER.
• Select
Active
if the current output is to be switched to
active. Select
Passive
if the current output is to be
switched to passive.
• Press ENTER.
Options\Channel A\Outputs\...\Output info
Options\Channel A\Outputs\...\Test signal
Options\Channel A\Outputs\...\Enter test value
Options\Channel A\Outputs\...\Test mea. range
Options\Channel A\Outputs\...\Enter test value
Special functions
Outputs
Active
Passive
Special functions\Outputs
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......