13 User modes
FLUXUS WD
13.1 StandardUser mode
105
UMFLUXUS_WDV1-3EN, 2018-07-16
• Enter the measured values of the transmitter.
• Press ENTER.
• Enter the measured values of the reference flowmeter.
• Press ENTER.
• Repeat the input for all pairs of measured values.
• Press ENTER after each input.
• Select
Yes
to also apply the calibration curve for negative flow velocities. Select
No
if it is not to be used for negative flow
velocities.
Start in measuring mode
For some application it is necessary to start the measurement in a particular measuring mode.
• Select the menu item
Special settings
in the program branch
Options
.
• Press ENTER until the menu item
Start in meas. mode
is displayed.
The menu item
Start in meas. mode
will only be displayed if FastFood or NoiseTrek mode is enabled.
• Select
TransitTime
,
FastFood
,
NoiseTrek
to start the measurement in the corresponding mode.
• Press ENTER.
• Select
Yes
to keep always the same the measuring mode. Select
No
to select another measuring mode by pressing key
during the measurement.
The measuring modes
FastFood
and
NoiseTrek
can only be selected if they are available and activated in the transmitter.
Options\Special settings\Multi-point calibration\Point x=act. value
Options\Special settings\Multi-point calibration\Point x=set value
Options\Special settings\Multi-point calibration\Bidirectional use
Options\Special settings\Start in meas. mode
Options\Special settings\Start in meas. mode\Only ... mode
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......