![Flexim FLUXUS WD Operating Instruction Download Page 91](http://html1.mh-extra.com/html/flexim/fluxus-wd/fluxus-wd_operating-instruction_2304736091.webp)
9 Measurement
FLUXUS WD
9.5 Display of parameters
91
UMFLUXUS_WDV1-3EN, 2018-07-16
Transducer temperature
If the transmitter is operating in the
SuperUser
or
SuperUser ext.
mode, the transducer temperature can be displayed
during the measurement.
By pressing key
during the measurement, it is possible to scroll to the transducer temperature display.
9.5
Display of parameters
The parameters can be displayed during the measurement.
• Press key
during the measurement.
The following display appears:
The measurement is running in the background. The symbol
is displayed in the status indication, see Fig. 9.6.
• Select
Show parameters
in the scroll list.
• Press ENTER.
• The program branch
Measurement
is displayed.
• Select a program branch to display the parameters.
The symbol
remains visible in all menus. All parameters and settings can be displayed.
Notice!
Never change the transducer distance during the measurement.
Fig. 9.5:
Display of the transducer temperature
1 – transducer temperature
Notice!
If the compliance of the specified transducer temperature has to be monitored, an event trigger can be set on the
temperature value.
Fig. 9.6:
Scroll list in the program branch
Measurement
1 – status indication
Measurement\Show parameters
Notice!
The parameters cannot be changed during the measurement. When attempting to change the parameters, the mes-
sage
Read-only mode
will be displayed.
The measurement has to be stopped in order to change the parameters.
1
1
Summary of Contents for FLUXUS WD
Page 1: ...FLUXUS WD UMFLUXUS_WDV1 3EN Operating instruction...
Page 6: ...Table of contents FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 6...
Page 150: ...18 Settings 18 9 Key lock FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 150...
Page 172: ...Annex D Conformity declarations FLUXUS WD 2018 07 16 UMFLUXUS_WDV1 3EN 172...
Page 173: ......
Page 174: ......