9 Measurement
FLUXUS G722ST-HT
9.2 Measurement settings
93
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
• Enter a test value for the selected physical quantity. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
9.2.6.2
Output of a status value
• Select the list item
Options\Outputs\...\Status
.
• Press ENTER.
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If
Other range
is selected, enter the values
Output MIN
and
Output MAX
.
The output range has to be > 10 % of the max. output value (
Output MAX
). An error message will be displayed if the
output range is smaller. The next possible value will be displayed.
Status OK
The status of the output signal is defined which is to be output if the measured value is valid.
• Select in the scroll list the value for
Status OK
.
• Press ENTER.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
Options\Outputs\...\Enter test value
Source item
Status
Output range
Status OK
Output info
Test signal
Test mea. range
see annex A
Options\Outputs\...\ Output range
Options\Outputs\...\Status OK
Options\Outputs\...\Output info