9 Measurement
9.1 Parameter input
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
82
9.1
Parameter input
The pipe and fluid parameters are entered for the selected
measuring point. The parameter ranges are limited by the
technical characteristics of the transducers and the
transmitter.
• Select the program branch
Parameters
.
• Press ENTER.
9.1.1
Transducer selection
• The transducer (here:
CDP2E52
) connected to the transmitter is displayed.
• Press ENTER.
This display will only be indicated if the transducers and the transducer module are connected to the transmitter.
The parameters can also be entered without connected transducers or transducer module.
• Select the list item
Connected transd.
• Press ENTER.
The display
Transducer not found
will be indicated if neither transducers nor transducer module are connected.
• Press ENTER.
Notice!
Avoid a simultaneous parameter entry via the keyboard and the USB, LAN or process interface. The parameter
records received via these interfaces will overwrite the current transmitter parameterization.
Parameters
Transducer
Connected transd.
Select transducer
Outer diameter
Pipe material
Pipe wall thickness
Lining
Roughness
Fluid
Fluid temp.
Extension cable
see annex A
Notice!
The transducers have to be selected depending on the application parameters (see technical specification).
Parameters\Clamp-on transducer CDP2E52
Parameters\Connected transd.