16 Data logger
16.1 Configuration of the data logger
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
128
16
Data logger
The transmitter has a data logger which stores the
measured values during the measurement.
The following data can be stored:
• date
• time
• measuring point number
• pipe parameters
• fluid parameters
• transducer data
• physical quantity
• unit of measurement
• measured values
Measured values transmitted via the outputs are also
stored in the data logger.
If the pulse values are transmitted via an output, the
corresponding flow quantity and the totalizer value are
stored in the data logger. In case of absolute pulse values,
the values of both totalizers are stored.
16.1
Configuration of the data logger
Starting time
It is possible to set a starting time if it is necessary to synchronize the storing of measured values for several transmitters.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Start storing
is displayed.
• Select the moment at which the storing has to start.
Special functions
Data logger
Configuration
Start storing
Ringbuffer
On
Off
Storage mode
Average
Sample
Store totalizers
Store diagnost. values
Store transd. temp.
Beep on storage
see annex A
Notice!
In order to store measured data, the data logger has
to be configured.
Special functions\Data logger\Configuration\Start storing
display
description
Immediately
The storing starts immediately.
Full 5 minutes
The storing starts in the next full 5 minutes.
Full 10 minutes
The storing starts in the next full 10 minutes.
Full 15 minutes
The storing starts in the next full 15 minutes.
Full 30 minutes
The storing starts in the next full 30 minutes.
Full hour
The storing starts in the next full 60 minutes.
Event-based
The storing starts when a defined event occurs.