14 Digital outputs
14.2 Configuration of a digital output as pulse output
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
122
Output function test
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select a test value.
• Press ENTER.
Options\Outputs\...\B1 Test signal
Options\Outputs\...\B1 Enter test value
Tab. 14.6: Output function test – signal
output mode
test value
description
Pulses per unit
The entered test value has to be
within the output range.
If the external measuring instrument displays the entered
value, the output functions correctly.
Pulse value
NO
• pulse output is de-energized
• measured value has to be high ohmic
NC
• pulse output is energized
• measured value has to be low ohmic