6 Mounting
6.2 Transducers
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
46
• Tighten the tensioning screws.
• Fix the spacing elements on the rail to mark the transducer position.
• Check whether the transducer distance of transducer pair B corresponds to the value displayed by the transmitter.
• Use a cable tie to fix the transducer cables in order to protect them from mechanical strain.
• Put the cover on the rail (see Fig. 6.44).
• Tighten the screws on both sides of the cover.
Remove the cover from the mounted transducer mounting fixture Variofix C as follows:
• Use a lever tool to remove the cover.
• Insert the lever tool in one of the 4 openings of the cover (see Fig. 6.45).
• Press the lever tool against the fixture.
• Bend the cover outwards and release it from the anchoring.
• Repeat the steps for the other 3 openings.
• Remove the cover from the rail.
Fig. 6.43: Transducers in the rail
1 – equipotential bonding terminal
2 – spring clip
3 – spacing element
4 – tensioning screw
5 – cable tie
Δx – measuring barrier distance
Fig. 6.44: Variofix C with transducers on the pipe
1 – screw
2 – cover
4
2
1
Δx
3
4
5
1
2