3 General principles
3.4 Undisturbed flow profile
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
16
Tab. 3.1:
Recommended distance from disturbances
D – nominal pipe diameter at the measuring point
l – recommended distance between disturbance and transducer position
disturbance: 90° elbow
inlet: l ≥ 10 D
outlet: l ≥ 5 D
disturbance: 2 × 90° elbow on different planes
inlet: l ≥ 40 D
outlet: l ≥ 5 D
disturbance: valve
inlet: l ≥ 40 D
outlet: l ≥ 5 D
disturbance: reducer
inlet: l ≥ 10 D
outlet: l ≥ 5 D
disturbance: compressor
inlet: l ≥ 20 D