14 Digital outputs
FLUXUS G722ST-HT
14.2 Configuration of a digital output as pulse output
121
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
• Enter the pulse width.
The range of possible pulse widths depends on the specification of the instrument (e.g., counter, PLC) that is to be
connected to the output.
• Press ENTER.
14.2.2
Pulse output by defining pulses per unit
• Select the list item
Pulses per unit
.
• Press ENTER.
• Select a list item:
– 0...1 kHz
– 0...5 kHz
– Other range
• Press ENTER.
If
Other range
is selected, enter a value for
Output MAX
.
• Enter a value for the number of pulses per unit.
• Press ENTER.
The unit of measurement will be displayed according to the actual physical quantity.
14.2.3
Output options
• Select the value for the idle state:
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
Options\Outputs\Digital output B1\...\Pulse width
Options\Outputs\Digital output B1\...\Pulse output\Pulses per unit
Options\Outputs\Digital output B1\...\B1 Output range
Options\Outputs\Digital output B1\...\Pulses per unit
Options\Outputs\Digital output B1\Source item\Pulse\...\Idle state
setting
description
NO
The pulse output is energized if a pulse is emitted and de-energized if no pulse is emitted (idle state).
NC
The pulse output is de-energized if a pulse is emitted and energized if no pulse is emitted (idle state).
Options\Outputs\Digital output B1\...\Output info