7 Connection
FLUXUS G722ST-HT
7.5 Temperature probe
71
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
7.5.2
Connection with extension cable
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part, the basic part and the reducer.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the connectors of the extension cable and the temperature probe.
• Connect the extension cable to the terminals of the transmitter.
7.5.3
Connection with junction box
Connection of the extension cable to the transmitter
• Remove the blind plug for the connection of the temperature probe.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut, the compression part, the basic part and the reducer.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter.
Connection of the extension cable to the junction box
• Remove the blind plug for the connection of the extension cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the extension cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the junction box.
Tab. 7.14: Terminal assignment (transmitter)
terminal
extension cable
T1a...T4a
red
T1A...T4A
gray
T1b...T4b
blue
T1B...T4B
white