13 User modes
FLUXUS G722ST-HT
13.2 ExpertUser mode
113
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
13.2.3
Profile correction
It is possible to select the following versions for the calculation of the fluid mechanics calibration factor k
Re
:
•
kRe 1.0
: profile correction (previous version)
•
kRe 2.0
: improved profile correction (current version, default)
The following steps are necessary to set the profile correction:
Selection of the version
• Select the menu item
Measurement settings
in the program branch
Special functions
.
• Press ENTER until the menu item
Profile correction
is displayed.
• Select a list item (default:
kRe 2.0
).
• Press ENTER.
Measuring point specific profile correction
In special cases a measuring point specific profile correction can be used.
• Select the program branch
Options
.
• Press ENTER.
• Select a channel for which the profile correction is to be set.
• Select the menu item
Special settings
.
• Press ENTER until the menu item
Profile correction
is displayed.
• Select
Off
to deactivate the profile correction for the channel.
• Select
Default
to use the global setting from the program branch
Special functions
for the profile correction.
• Select
Customized
to use a measuring point specific profile correction.
• Press ENTER.
In case the list item
Customized
has been selected, the parameters of the measuring point specific profile correction will
now be displayed. The parameters of the profile correction are preferably transmitted to the transmitter via the service
interface but can also be entered here.
Special functions\Measurement\Measurement settings\Profile correction
Options\Special settings\Profile correction