6 Mounting
FLUXUS G722ST-HT
6.2 Transducers
35
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.17).
• Tighten the tension strap.
• Tighten the clasp screw.
Ratchet clasp
• Insert the tension strap into its clamps and the metal spring (see Fig. 6.19).
• Position the ratchet clasp, the tension strap clamps and the metal spring on the pipe (see Fig. 6.15):
– On a horizontal pipe, mount the tension strap clamps laterally to the pipe, if possible.
– Mount the metal spring opposite the ratchet clasp.
• Make sure that the 2 tension strap clamps are as exactly as possible opposite each other. To do this, check whether the
distance between the tension strap clamps is half the pipe circumference.
• Place the tension strap around the pipe and pass it through the slot of the clasp screw (see Fig. 6.20).
• Tighten the tension strap.
• Cut off the protruding tension strap (see Fig. 6.20).
• Tighten the clasp screw.
Fig. 6.18: Tension strap with band clamp clasp
1 – clasp screw
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Notice!
In order to release the screw and the tension strap, press the lever down (see Fig. 6.20).
Fig. 6.19: Tension strap with metal spring and clamp
1 – metal spring
2 – tension strap clamp