6 Mounting
FLUXUS G722ST-HT
6.2 Transducers
43
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
• Insert approx. 100 mm of the tension strap into one of the slots of the tension strap clamp and bend it (see Fig. 6.35).
• Insert the long end of the tension strap into the metal spring and the 2 tension strap clamps (see Fig. 6.36).
Place the tension strap around the pipe (see Fig. 6.37).
• Position the tension strap clamps and the metal spring on the pipe:
– On a horizontal pipe, mount the tension strap clamps laterally to the pipe, if possible.
– Mount the metal spring on the bottom side of the pipe.
• Make sure that the 2 tension strap clamps are as exactly as possible opposite each other. To do this, check whether the
distance between the tension strap clamps is half the pipe circumference.
• Insert the long end of the tension strap into the second slot of the tension strap clamp (see Fig. 6.37 a).
• Tighten the tension strap and bend it.
• Bend both ends of the tension strap (see Fig. 6.37 b).
• Repeat the steps for the second tension strap.
• Position the tension straps at the distance s (see Fig. 6.37).
• Put the rails on the tension strap clamps.
• Fix the rails to the tension strap clamps using the screws (see Fig. 6.38).
• Tighten the screws.
Fig. 6.35: Tension strap with clamp
Fig. 6.36: Tension strap with metal spring and clamp
1 – tension strap clamp
1 – tension strap clamp
2 – metal spring
Fig. 6.37: Tension strap with metal spring and clamps on the pipe
1 – tension strap clamp
s = length of the rail - 33 mm
Fig. 6.38: Rail on the pipe
1 – screws
1
1
2
1
a
b
s
1
1