![Flexim FLUXUS G722ST-HT Operating Instruction Download Page 37](http://html1.mh-extra.com/html/flexim/fluxus-g722st-ht/fluxus-g722st-ht_operating-instruction_2304717037.webp)
6 Mounting
FLUXUS G722ST-HT
6.2 Transducers
37
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
Band clamp clasp
• Insert the tension strap into its clamp (2) (see Fig. 6.23).
• Place the tension strap around the pipe and insert it into the clasp and the tension strap clamp on the opposite side
(see Fig. 6.24).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nuts of the tension strap clamps (2), but not too firmly in order not to damage the tension strap
(see Fig. 6.23).
Quick release clasp
• Insert the tension strap into its clamp (2) and the metal spring (see Fig. 6.23 and Fig. 6.25).
• Place the tension strap around the pipe and insert it into the clasp and the tension strap clamp on the opposite side.
• Position the metal spring opposite the clasp.
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nuts of the tension strap clamps (2), but not too firmly in order not to damage the tension strap
(see Fig. 6.23).
Fig. 6.23: Rail on the pipe
1 – tension strap clamp
2 – tension strap clamp
3 – metal spring
4 – nut
5 – clasp
Fig. 6.24: Tension strap with band clamp clasp
Fig. 6.25: Tension strap with quick release clasp
and metal spring
1 – clasp screw
1 – clasp screw
2 – metal spring
3
4
2
5
1
1
2
1