![Flexim FLUXUS G722ST-HT Operating Instruction Download Page 112](http://html1.mh-extra.com/html/flexim/fluxus-g722st-ht/fluxus-g722st-ht_operating-instruction_2304717112.webp)
13 User modes
13.2 ExpertUser mode
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
112
• Enter a limit of the flow velocity for the measurement in flow direction.
• Press ENTER.
If the flow velocity is higher than this limit, it will be marked as invalid. The physical quantity cannot be determined.
UNDEF
will be displayed.
• Enter a limit of the flow velocity for the measurement against the flow direction.
• Press ENTER.
If the flow velocity is lower than this limit, it will be marked as invalid. The physical quantity cannot be determined.
UNDEF
will be displayed.
13.2.2
Threshold value for valid measurement
• Select the menu item
Special settings
in the program branch
Options
.
• Press ENTER until the menu item
Crest factor threshold
is displayed.
• Enter the threshold value (default: 6).
• Press ENTER.
Options\Special seFlow velocity limit
Options\Special settings\-Flow velocity limit
Fig. 13.1: Flow velocity outside the valid range
1 – physical quantity
2 – flow velocity
Notice!
If the limit of the flow velocity
+Flow velocity limit
is too low or
-Flow velocity limit
is too high,
a measurement might be impossible because most of the measured values will be marked as invalid.
Options\Special settings\Crest factor threshold
Notice!
A reduced threshold value allows for measurements under difficult conditions, but increases the risk of measurement
errors.
1