14 Digital outputs
14.1 Configuration of a digital output as binary output
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
116
14
Digital outputs
If the transmitter is equipped with outputs, they have to be
configured. For the configuration of the analog output, see
section 9.2.6.
In addition to the analog outputs, the transmitter can be
equipped with digital outputs. A digital output combines the
functions of the following outputs:
• binary output (output of binary switching conditions)
• pulse output (integrating output of quantities)
• frequency output (scaled output of flow quantities)
These functions depend on the selected physical quantity.
14.1
Configuration of a digital output as binary output
A digital output switches if one of the following switching conditions is met:
– the measured value exceeds or falls below a limit
– the measured value lays within or outside a defined range
– a measurement is not possible
– an event occurs
Assignment of an output
• Select the menu item
Outputs
.
• Press ENTER.
• Select the output (here:
Digital output B1
).
• Press ENTER.
If the output has already been activated, it is displayed as follows:
Digital output B1 (
)
.
Options
Outputs
Select an output
Enable the output
Source item
see annex A
Tab. 14.1: Output via digital outputs
source item
binary output
pulse output
frequency
output
status value
event value
physical quantities
Flow quantities
x
x
Totalizers
x
x
Pulse
x
Fluid properties
x
x
Diagnostic values
(Crest factor)
x
Miscellaneous
(Custom. Input 1...4)
x
x
events
Event trigger
x
Options\Outputs
Options\Outputs\Digital output B1(--)