9 Measurement
9.2 Measurement settings
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
90
Depending on the selected source item, it is possible to output measured values, status values or event values.
9.2.6.1
Output of a measured value
• Select the list item
Options\Outputs\...\Values
.
• Press ENTER.
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If
Other range
is selected, enter the values
Output MIN
and
Output MAX
.
The output range has to be > 10 % of the max. output value (
Output MAX
). An error message will be displayed if the
output range is smaller. The next possible value will be displayed.
An error value can be defined which is output if the source item cannot be measured.
• Select a list item for the error output.
• Press ENTER.
• If
Other value
is selected, enter an error value. The value has to be within the output range.
• Press ENTER.
Tab. 9.2:
Output of measured values, status values or event values
source item
measured value
event value
value
status
physical quantities
Flow quantities
x
x
Totalizers
x
x
Fluid properties
x
x
Miscellaneous
(Custom. Input 1...4)
x
x
Diagnostic values
(except
Transd. temp. violat.
)
x
events
Diagnostic values\
Transd. temp. violat.
x
Event trigger
x
Source item
Values
Output range
Error value
Measuring range
Output info
Test signal
Test mea. range
see annex A
Options\Outputs\...\Output range