14 Digital outputs
FLUXUS G722ST-HT
14.1 Configuration of a digital output as binary output
117
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
• Select
Yes
to change the settings for an already assigned output or to assign a new one.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
Assignment of a source item
Depending on the selected source item, status or event values can be output.
• Select the source item.
• Press ENTER.
• Select the list item
Status
.
• Press ENTER.
14.1.1
Definition of the switching function for the status/event value
• Select the switching function for the output of the status/event value.
• Press ENTER.
If no measurement is carried out, all binary outputs are open (de-energized), independent of the set switching condition.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
Options\Outputs\Digital output B1\Enable B1
Options\Outputs\...\Source item
Tab. 14.2: Selection of the switching condition
property
switching function
description
Status OK
(status value)
NC
• valid measured value: binary output is closed
• invalid measured value: binary output is open
NO
• valid measured value: binary output is open
• invalid measured value: binary output is closed
Idle state
(event value)
NO
• event occurs: binary output is closed
• event has not occurred yet: binary output is open
NC
• event occurs: binary output is open
• event has not occurred yet: binary output is closed
Options\Outputs\...\Output info