![Fisher-Rosemount 3095FT Manual Download Page 68](http://html1.mh-extra.com/html/fisher-rosemount/3095ft/3095ft_manual_2300591068.webp)
Rosemount Model 3095FT HART Flow Transmitter
4-20
Configuration
Flow
Gas Properties (Detail)
The displayed gas properties screen is determined by the calculation
method selected in the Flow Parameters screen. If “AGA8 Detail” is
selected, then the AGA 8 (1992) Full Composition Analysis screen is
displayed (Figure 4-8a). This method allows entry of up to 21 different
gas composition mole percentages. Listed below is the variable name for
each abbreviation. Table 4-1 on Page 4-16 identifies the valid range for
each variable.
N2
Nitrogen
iC5
i-Pentane
CO2
Carbon Dioxide
nC5 n-Pentane
H2s
Hydrogen Sulfide
C6
Hexane
H20
Water
C7
Heptane
He
Helium
C8
Octane
C1
Methane
C9
Nonane
C2
Ethane
C10 Decane
C3
Propane
O2
Oxygen
iC4
i-Butane
CO
Carbon Monoxide
nC4
n-Butane
H2
Hydrogen
Argon Argon
When entering numbers into the detail gas properties screen, the Total
Mole % field indicates the sum of all percentages entered. The Total
Mole % field must add up to 100.0000 percent for the User Interface to
accept the new values.
To zero all 21 fields, select clear. The normalize button provides a
method to automatically modify all non-zero values so that they add up
to 100.0000.
After the mole percentages are entered, select Ok to display the AGA 8
Gas Properties screen (Figure 4-8).
30
95
-30
950
03
8
FIGURE 4-8a. Full Composition
Analysis Screen (AGA8-Detail).
Summary of Contents for 3095FT
Page 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Page 2: ......
Page 4: ......
Page 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Page 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Page 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Page 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Page 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Page 149: ...A 3 Appendix A...
Page 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Page 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Page 153: ...B 3 Appendix B...
Page 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Page 155: ...B 5 Appendix B...
Page 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Page 157: ...B 7 Appendix B...
Page 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Page 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Page 161: ...C 3 Appendix C...
Page 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Page 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Page 165: ...D 3 Appendix D...
Page 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Page 167: ...D 5 Appendix D...
Page 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Page 183: ......