Rosemount Model 3095FT HART Flow Transmitter
4-14
USER INTERFACE
SOFTWARE SCREENS
This section illustrates each major Model 3095FT User Interface
screen, and provides information on using the screen.
Configuration Screens
The configuration screens define a configuration that can be sent to a
transmitter (Send Xmtr), and/or saved to a file (Save Config File).
NOTE
Security Level SL1 provides read-only access. Security Level SL2 and
SL3 provides read/write capability to an attached transmitter.
Configuration
Flow
Flow Parameters
The flow parameters screen (Figure 4-5) should be completed first
whenever preparing a Model 3095FT configuration. This screen
defines:
Software Tag
The software tag uniquely identifies the transmitter. This tag is also
used to identify audit trail data retrieved from the transmitter.
Meter Tube Bore @ 68F (in)
Identifies the diameter of the meter tube bore.
Meter Tube Matl
Allows selecting one of five meter tube materials: carbon steel, SST 304,
SST 316, Hastelloy C, and Monel.
Orifice Bore @ 68F (in)
Identifies the diameter of the orifice bore.
NOTE
The Orifice Bore/Meter Tube Bore ratio (
β
) should be within the
following range:
Flange Tap: 0.1
≤ β ≤
0.75
Pipe Tap: 0.2
≤ β ≤
0.67
Orifice Plate Matl
Allows selecting one of five orifice plate materials: carbon steel, SST
304, SST 316, Hastelloy C, and Monel.
Orifice Tap Type
Allows selecting either flange or pipe taps.
FIGURE 4-5. Flow Parameters Screen.
30
95
-30
950
00
1
Summary of Contents for 3095FT
Page 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Page 2: ......
Page 4: ......
Page 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Page 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Page 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Page 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Page 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Page 149: ...A 3 Appendix A...
Page 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Page 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Page 153: ...B 3 Appendix B...
Page 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Page 155: ...B 5 Appendix B...
Page 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Page 157: ...B 7 Appendix B...
Page 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Page 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Page 161: ...C 3 Appendix C...
Page 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Page 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Page 165: ...D 3 Appendix D...
Page 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Page 167: ...D 5 Appendix D...
Page 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Page 183: ......