![Fisher-Rosemount 3095FT Manual Download Page 134](http://html1.mh-extra.com/html/fisher-rosemount/3095ft/3095ft_manual_2300591134.webp)
Rosemount Model 3095FT HART Flow Transmitter
7-12
FLAT COVER
CONNECTION
ASSEMBLIES
The Flat Cover Connection Head (P/N 00079-0325-0002) is for general
purpose and spring-loaded sensors. The terminal block has six
terminals for either single- or dual-element sensors.
Sensor Connections
½–14 NPT mounting thread. Screw terminals for lead wire connections.
Electrical Connection
¾–14 NPT conduit.
Materials of Construction
Housing:
Low-copper aluminum alloy.
O-ring seal:
Silicone rubber.
Terminals:
Nickel-plated brass.
Temperature Limits
-100 to 200 °C (-148 to 392 °F).
Hazardous Locations Certifications
(Flat Connection Head Assembly)
Explosion Proof
Approved by Factory Mutual (FM) for Class I, Division 1, Groups B,
C, and D; Class II, Division 1, Groups E, F, and G; and Class III,
Division 1, hazardous locations.
Approved by the Canadian Standards Association (CSA) for Class I,
Division 2, Groups A, B, C, and D; Class 1, Division 1, Groups C and
D; Class II, Division 1, Groups E, F, and G; Class III, Division 1 and
enclosure 4.
Weight
Flat cover type: 1 lb 9 oz (0.71 kg).
FIGURE 7-1. Connection Head
Dimensional Drawing.
S
e
nso
rs-
033
5B
,
0
035
C
FLAT COVER
CONNECTION HEAD
WITH COVER REMOVED
3.5 (88.9)
NOTE
Dimensions are in inches (millimeters).
½–14 NPT
Terminal Block
(Six Terminals)
4.9
(125)
TERMINAL VIEW
¾–14 NPT
Summary of Contents for 3095FT
Page 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Page 2: ......
Page 4: ......
Page 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Page 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Page 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Page 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Page 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Page 149: ...A 3 Appendix A...
Page 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Page 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Page 153: ...B 3 Appendix B...
Page 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Page 155: ...B 5 Appendix B...
Page 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Page 157: ...B 7 Appendix B...
Page 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Page 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Page 161: ...C 3 Appendix C...
Page 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Page 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Page 165: ...D 3 Appendix D...
Page 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Page 167: ...D 5 Appendix D...
Page 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Page 183: ......