5-13
Troubleshooting and Maintenance
Removing the Sensor from
the Electrical Housing
6. Loosen the housing rotation set screw with a 5/64-inch hex
wrench and back off one full turn.
7. Unscrew the housing from the module, making sure the shroud
and sensor cable do not catch on the housing. Damage can occur
to the cable if the internal shroud and sensor cable rotate with
the housing. Carefully pull the shroud and sensor ribbon cable
assembly through the housing opening.
The sensing module is a complete assembly and cannot be further
disassembled.
Before removing the sensor module from the electrical
housing, disconnect the electronics board power cable from
the sensor module. This will prevent damage to the sensor
module ribbon cable.
If the Coplanar flange has been removed, take care not to
damage the isolating diaphragm after disassembly. Damage
to the isolating diaphragm may lead to inaccurate
measurements.
Summary of Contents for 3095FT
Page 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Page 2: ......
Page 4: ......
Page 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Page 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Page 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Page 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Page 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Page 149: ...A 3 Appendix A...
Page 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Page 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Page 153: ...B 3 Appendix B...
Page 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Page 155: ...B 5 Appendix B...
Page 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Page 157: ...B 7 Appendix B...
Page 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Page 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Page 161: ...C 3 Appendix C...
Page 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Page 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Page 165: ...D 3 Appendix D...
Page 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Page 167: ...D 5 Appendix D...
Page 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Page 183: ......