4-13
Using the Model 3095FT User Interface Software
Bench Calibration
Procedure
After a transmitter is bench configured, the transmitter can be bench
calibrated.
1. Enter either the system administrator or a maintenance level
password.
2. Set or verify the Damping settings.
3. Place the transmitter into Off-line or On-line Maintenance.
4. Perform sensor trim procedures:
NOTE
Although the Model 3095FT can calculate and log static pressure as
absolute or gage, flow is always calculated using the absolute pressure
sensor measurement.
a.Trim SP Offset (zero).
b.Trim SP Slope (span).
c. Trim DP Offset (zero).
d.Trim DP Slope (span).
e. Trim PT Offset (zero).
f. Trim PT Slope (span).
Field Calibration
Procedure
To correct for mounting position effects, field calibrate the Model
3095FT after installation:
1. Establish communications (see Page 4-4)
2. Enter either the system administrator or a maintenance level
password.
3. Place the transmitter into Off-line or On-line Maintenance.
4. Perform a Trim DP Offset (zero).
5. (Optional) If a barometer that is four times as accurate as the
Model 3095FT AP sensor is available, perform an AP Offset.
Collect New Log Data
During this procedure, the Model 3095FT User Interface checks the
archive, and only retrieves data that it does not yet have.
1. Establish communications (see Page 4-4)
2. Enter either the system administrator, a maintenance level
password, or an operator password.
3. Click
in the tool bar.
Collect All Log Data
This procedure retrieves all data from the connected transmitter.
1. Establish communications (see Page 4-4)
2. Enter either the system administrator, a maintenance level
password, or an operator password.
3. Click
in the tool bar.
Collect Log Data per
System Defaults
During this procedure, the Model 3095FT User Interface collects log
data according to the defaults set in the Sys Admin, Options screen.
1. Establish communications (see Page 4-4)
2. Enter either the system administrator, a maintenance level
password, or an operator password.
3. Click
in the tool bar.
Summary of Contents for 3095FT
Page 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Page 2: ......
Page 4: ......
Page 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Page 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Page 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Page 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Page 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Page 149: ...A 3 Appendix A...
Page 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Page 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Page 153: ...B 3 Appendix B...
Page 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Page 155: ...B 5 Appendix B...
Page 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Page 157: ...B 7 Appendix B...
Page 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Page 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Page 161: ...C 3 Appendix C...
Page 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Page 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Page 165: ...D 3 Appendix D...
Page 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Page 167: ...D 5 Appendix D...
Page 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Page 183: ......