73
www.fi-measurement.com
W
Reference mass (100%)
Percentag
e
9
X
Percentage value
Percentag
e
10
Y
Weight Hold (Animal weighing)
Animal
Weighing
9
Note A
: - 10 for counting mode; 9 for all other modes.
Note B
: - 6 for ACCEPT; 11 digits for BELOW LIMIT and ABOVE LIMIT.
Note C
: - Good for all except animal weighing mode.
B.B.2 Data Format
64
65
66
Protocol of data answered by this instrument is illustrated on below table.
Commands Data example and format Description
J
ST,NT 123.567,123.567,kg
•
2 digits motion status (ST = stable; US = unstable)
•
1 digit comma separation
•
2 digits Net/Gross ( NT = Net; GS =Gross)
•
1 comma separation
•
1 digit sign. Positive = space. Negative = minus (-)
•
7 digits weight value including decimal point. If there
is no decimal point, then the first character = space
•
1 digit comma separation
•
7 digits tare weight including decimal point
•
1 digit comma separation
•
2 digits weight unit
K
22.06.2012
•
8-digit date format depends on F4 setting
64
All data = align to right.
65
Insignificant figures = space.
66
All data end up with CR, LF.