27
www.fi-measurement.com
6.4.3 Repeated Tare (F13)
9
When F13 is set to OFF, this instrument does not permit multiple tare
operation. Tare effect can only be cancelled when container is removed and
gross weight = zero.
When F13 is set to ON, this instrument will permit multiple tare operation
provided that both of the below requirements are met: -
a. The tare operation does not permit a reduction of the value of the tare;
b. The tare effect can only be cancelled when there is no load on the
platform.
6.4.4 Preset Tare (F63)
10
A pre-determined tare weight can be entered via keyboard. To enable this
function, set F63 to ON, then set also F12 to OFF.
During normal operation, press
[0]
then followed by the pre-determined tare
weight through numeric keys and press
[M+]
to enter.
To cancel preset tare effect, remove all loads from platform then press
[TARE]
.
Notes: -
1.
The pre-determined tare weight entered will be rounded to the nearest
division of the instrument. This does not affect the accuracy of the
subsequent weighing and operation.
2.
Refer to
Specifications
for maximum tare range.
3.
Manual tare is possible when preset tare is in function.
4.
Preset Tare is also governed by Repeated Tare (F13)
9
Set F13 = ON to enable Repeated Tare Function.
10
This is not a legal for trade function.
Summary of Contents for AFM18
Page 2: ...1 www fi measurement com ...
Page 12: ...11 www fi measurement com 3 2 Connection Points AFM18 FM18 FM18S ...
Page 20: ...19 www fi measurement com 5 4 Internal Function Table ...
Page 21: ...20 www fi measurement com ...
Page 22: ...21 www fi measurement com ...
Page 23: ...22 www fi measurement com ...
Page 24: ...23 www fi measurement com ...
Page 25: ...24 www fi measurement com ...