![Fidelity Measurement AFM18 Operation Manual Download Page 48](http://html.mh-extra.com/html/fidelity-measurement/afm18/afm18_operation-manual_2287074048.webp)
47
www.fi-measurement.com
6.16.2 Constant Dispensing Sequence Diagram
Refer to below illustration diagram for dispensing sequence.
6.16.3 Constant Dispensing Sequence Description
Seq.
Point
Description
1
C
a.
[FUNC]
key is pressed.
b.
Instrument displays
Start
.
2
C
D
Time delay as set forth in Delay 1.
3
D
a.
Dispensing sequence starts.
b.
Relay 2 (SP1) starts action.
c.
Fast dispensing starts.
4
E
a.
Weight value of SP1 is reached.
b.
Relay 2 stops action.
c.
Fast dispensing stops.
5
E
F
a.
Time delay as set forth in Delay 2. This is the
stabilization / waiting time before SP2 (slow
dispensing) starts.
b.
If a weight value
less
than SP1 is detected in this
period, Seq.
3
and
4
will be repeated