4.3 Connecting Other Devices
6
4.3.1 Connection with Weighing Platform (Load Cell)
Connect this instrument with a weighing platform (load cell) through
LOAD
CELL CONNECTOR
located at the back according to the below pin
assignment table.
If a 4-wire load cell or junction box is used, short-circuit pin 1 & 2 and pin 3 & 4.
Otherwise, this instrument will not work.
LOAD CELL
CONNECTOR
PIN #
ASSIGNMENT
1
EXCIve
2
SENSE +ve
3
EXCITATION
–ve
4
SENSE
–ve
5
ve
6
SIGNAL
–ve
7
GROUND
4.3.2. Connecting RS232 to Computer
RS232 COMPORT ON
INSTRUMENT
COM PORT ON COMPUTER
(DB9)
(DB25)
2 = RXD
3 = TXD
3 = TXD
3 = TXD
2 = RXD
2 = RXD
5 = GND
5 = GND
7 = GND
6
Turn this instrument off and cut off power before making any connections or disconnections.