Multi-zone ART Workstation User Manual Rev. 6.0
95
8.
N
2
gas outlet with 0.6
–
1.0 atm above ambient if the clinic uses reduced oxygen
levels.
9.
Tubes that fit 4 mm hose nipple and HEPA filter.
10.
Access to a PC with USB for the data logging.
34.3 Preparing for installation
•
Bring the
"Installation report"
form. Make sure it is the latest and current version
only.
•
Fill out the following blank boxes in the form: the Multi-zone ART Workstation
serial number (S/N) and customer.
•
The service tool kit is checked for content before every installation trip to ensure
it contains the necessary tools.
•
Always bring the latest versions of firmware and data logging software. Bring
these files on a labeled memory stick to the service site.
34.4 Bring the following to the installation site
•
"Installation report" form.
•
Service manual for the Multi-zone ART Workstation.
•
Updated service tool kit.
•
Memory stick with the latest released firmware & software.
•
High precision thermometer with a resolution not less than 0.1 °C.
•
Calibrated gas analyzer with precision at least 0.1% for CO
2
and O
2
•
Extension cable for USB connection.
34.5 Installation procedure at the site
1.
Lift the cabinet using a stacker or forklift
Figure 34.1
Multi-zone ART Workstation, cabinet