Multi-zone ART Workstation User Manual Rev. 6.0
93
33.13 Internal 0.2µ filter for CO
2
gas
The round-shape in line 0.2µ HEPA filter for CO
2
gas removes any particles found in the
incoming gas. Failure to use the HEPA filter may cause damage to the high precision flow
sensor, which calculates the amount of gas entering the system and disturbing the gas
flow system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming CO
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
33.14 Internal 0.2µ filter for N
2
gas
The round-shape in line 0.2µ HEPA filter for N
2
gas removes any particles found in the
incoming gas. Failure to use the HEPA filter may cause damage to the high precision flow
sensor, which calculates the amount of gas entering the system and disturbing the gas
flow system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming N
2
gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
33.15 Firmware update
If Esco Medical has released a newer version of the firmware, this should be installed on
the Multi-zone ART Workstation during the yearly scheduled service.
Please refer to the service manual for instructions on how to update the firmware.