Multi-zone ART Workstation User Manual Rev. 6.0
73
Guidance and manufacturer’s declaration –
electromagnetic immunity
The Multi-zone ART Workstation
is intended for use in the electromagnetic environment specified below.
The customer or the Multi-zone ART Workstation
user should ensure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance
level
Electromagnetic environment-
guidance
Conducted RF IEC
61000-4-6
Radiated RF IEC
61000-4-3
10 Vrms 150kHz to 80 MHz in
ISM bands
3V/m 80 MHz to 2.5 GHz
3 V/m from
80MHz to 2.5
GHz
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the Multi-
zone ART Workstation, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the transmitter's frequency.
Recommended separation dis-
tance
d=0.35 P
d=0.35 P
80MHz to 800 MHz
d=0.7 P
800MHz to 2.5 GHz
P is the maximum output power
rating of the transmitter in watts
(W) according to the transmitter
manufacturer,
d
is the recom-
mended separation distance in
meters (m).
As determined by an electromag-
netic site survey, field strengths
from fixed RF transmitters should
be less than the compliance level
in each frequency range.
Interference may occur near
equipment marked with.