Multi-zone ART Workstation User Manual Rev. 6.0
67
The 3 action buttons are located on the right-hand side:
•
The “
R
econnect”
button allows the system to reconnect with the sensors (in case
of data loss due to USB connection issues).
•
The “
M
inimize”
button toggles the full-screen format off for the Workstation Log-
ger.
•
The “
E
xit”
button shuts down the Workstation Logger software.
No data will be stored and all safety surveillance functions will be operative if
the Workstation Logger is shut down.
The AIO PC can easily be used to display an image from any USB-type microscope camera.
A standard USB socket is located under the humidification system. It can be used to load
the microscope camera drivers or other software onto the AIO PC. When the software is
set up, the microscope camera can be connected to the USB socket and the image shown
on the screen.
A force function is built into the Workstation Logger. When the microscope
camera image uses the full screen on the AIO PC, the system will visually alert the
user to any alarm conditions and make a rapid shift to the Workstation Logger's
full view possible.
24 Maintenance
The Multi-zone ART Workstation
is designed to be easy to use, but the reliable and safe
operation of this equipment is based on the following conditions:
1.
Correct calibration of temperature and gas level, using high-precision equipment
in the intervals prescribed based on clinical practice at the laboratory, where
Multi-zone ART Workstation is used. The manufacturer recommends that the pe-
riod between validation should be no longer than 14 days.
2.
In-line HEPA filters must be replaced yearly during annual maintenance.
3.
According to the clinical practice intervals, suitable cleaning is in the laboratory
where Multi-zone ART Workstation is used. The manufacturer does not recom-
mend periods longer than 14 days between cleaning.
It is essential to perform the inspection and service at the intervals indicated
in the MAINTENANCE section below. Failure to do so can have serious adverse out-
comes, causing the unit to stop performing as expected and cause damage to sam-
ples, patients or users.