Multi-zone ART Workstation User Manual Rev. 6.0
18
6.
Switch on the Multi-zone ART Workstation
with the switch under the tabletop
(next to the mains cable).
7.
Power up the PC by pressing the button in the middle of the inner wall work area.
8.
Observe for standard functionality.
9.
Let the unit warm up and stabilize for 20 min.
10.
Follow the guidelines in the Validation guide.
11.
Complete user training and finish reading instructions.
12.
After a burn-in phase of 24-hours, the unit is ready to be used IF testing is success-
ful.
Clean and disinfect the device before use. It is not delivered sterile or in a
clinically acceptable clean state. Look through the cleaning instructions section in
this manual for the manufacturer's recommended guidelines!
10 Mains connection
The Multi-zone ART Workstation comes with a detachable mains power cord. The power
cord is prepared for the country in which the unit is intended to be used.
Do not defeat the safety purpose of the grounding-type plug! A grounding-type
plug has two blades and a prong, which is provided for your safety. If the provided
plug does not fit into your outlet, consult an electrician to replace the outlet.
The power requirement is 230V 50 Hz OR 115V 60Hz. The built-in power supply has a
switch-mode that automatically adjusts to the correct mains power between 100V-250V
AC 50-60 Hz.
Figure 10.1
Power supply
11 Gas connection and the humidification system
Under the SINGLE tabletop, one single gas input (black and blue) is located, whereas, un-
der the DUAL tabletop, two single gas inputs are located.