Multi-zone ART Workstation User Manual Rev. 6.0
40
The default O
2
setpoint is 5.0%.
To change the O
2
concentration setpoint, follow these instructions:
1.
When the display shows the O
2
concentration:
2.
Hold down the SP key and use (⇧) and (⇩) keys to adjust
the setpoint: one keypress
corresponds to a 0.1 change.
3.
After changing the temperature, let go of the SP key. The value is now stored.
If the display does not show the current O
2
reading, the
(⇨) key will toggle between the
temperature, CO
2
, O
2
and heated mode readings.
14.5 Heated Mode
The table plate has 4 heated modes.
To change heated mode, follow these instructions:
1.
When the display shows the current heated mode:
2.
Hold down the SP key and use (⇧) and (⇩) keys to adjust the
heated mode.
3.
After changing the temperature, let go of the SP key. The mode is now stored.