Multi-zone ART Workstation User Manual Rev. 6.0
43
temperature of a separate area.
Each of the 12 available areas is equipped with a separate temperature sensor and heater,
allowing the user to adjust the temperature in every area separately, thus achieving
higher precision.
Figure 15.1
PID controllers on the table plate
Each area can be calibrated separately, using the item corresponding to the respective
area in the menu. These items are placed in the menu and they are named: T1 CAL, T2
CAL, T3 CAL, T4 CAL, T5 CAL, T6 CAL, T7 CAL, T8 CAL, T9 CAL, T10 CAL, T11 CAL and T12
CAL.
To calibrate the temperature in a particular area, please find the corresponding sensor
name and adjust it according to a measurement taken using a high-precision
thermometer.
Esco Medical recommends only a suitable and calibrated device with a precision of at
least 0.1 °C.
Temperature calibration is done by adjusting the Tx (where x is the sensor
number) according to a measurement done on the spot relevant to the dish
placement.
After temperature adjustment, give it at least 15 minutes for the temperature
to stabilize, use the thermometer to verify the correct temperature on each area.
Tape the calibrated thermometer sensor to the middle of the zone. It may be