Multi-zone ART Workstation User Manual Rev. 6.0
81
PASS: The value must be 0.4
–
0.6 bar.
Please refer to the User manual sections for more information.
31.5 CO
2
gas pressure check (only for models with a built-in gas mixer)
The Multi-zone ART Workstation requires a pressure of 0.4
–
0.6 bar (5.80
–
8.70 PSI) on
the input gas line. Therefore, the gas pressure must be held stable at all times.
For safety, this unit has a built-in digital gas pressure sensor that monitors the incoming
gas pressure and alerts the User if any drop is detected.
Remove the inlet gas line for the CO
2
gas. Attach the gas line to the gas pressure measuring
device.
PASS: The value must be 0.4
–
0.6 bar.
Please refer to the User manual sections for more information.
31.6 N
2
gas pressure check (only for models with a built-in gas mixer)
The Multi-zone ART Workstation requires a pressure of 0.4
–
0.6 bar (5.80
–
8.70 PSI) on
the input gas line. Therefore, the gas pressure must be held stable at all times.
For safety, this unit has a built-in digital gas pressure sensor that monitors the incoming
gas pressure and alerts the User if any drop is detected.
Remove the inlet gas line for the N
2
gas. Attach the gas line to the gas pressure measuring
device.
PASS: The value must be 0.4
–
0.6 bar.
Please refer to the User manual sections for more information.
31.7 Voltage supply
The voltage on-site must be verified.
Measure the output plug on the UPS that the Multi-zone ART Workstation will be
connected to. Also, check that the UPS is attached to a properly grounded mains outlet.
Use a multimeter set for AC.