Multi-zone ART Workstation User Manual Rev. 6.0
87
The table below shows time intervals in which components must be replaced.
Table 33.1
Service interval plan
Component name
Every
year
Every 2
years
Every 3
years
Every 4
years
HEPA in-line filter for premixed gas
X
HEPA in-line filter for CO
2
gas
1
X
HEPA in-line filter for N
2
gas
1
X
O
2
sensor
1
X
CO
2
sensor
1
X
Gas box
2
X
Internal gas pump
X
Proportional valves
X
Gas lines
X
Flow sensors
X
Pressure regulators
X
Internal 0.2µ filter for premixed gas
X
Internal 0.2µ filter for CO
21
X
Internal 0.2µ filter for N
21
X
A firmware update (if a new version has been
released)
X
33.1 HEPA in-line filter for premixed gas
The round-shape in-line 0.2µ HEPA filter for premixed gas removes any particles found
in the incoming gas. Failure to use the HEPA filter may cause damage to the high precision
flow sensor, which calculates the amount of gas entering the system and disturbing the
gas flow system.
Please follow these safety precautions when changing the filter:
•
Always use the original filter (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the filter once every year.
•
Failure to change the filter on time will result in low/no cleaning of incoming pre-
mixed gas.
•
Warranty void if wrong/ non-original filter is used.
Please refer to the service manual for replacement instructions.
33.2 HEPA in-line filter for CO
2
gas
The round-shape in-line 0.2µ HEPA filter for CO
2
gas removes any particles found in the
1
Applicable only for Multi-zone ART Workstation with a built-in gas mixer
2
Applicable only for Multi-zone ART Workstation without a built-in gas mixer