Operation
Liquiline System CA80TN
64
Hauser
Behavior after power failure and other settings
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
After power failure
Selection
• Last mode
• Manual mode
Factory setting
Last mode
Setting specifying how the analyzer should
behave after a power failure, or when the power
is back on.
Last mode
: The analyzer remains in the mode
that was last set. Example: The automatic mode
was set. The analyzer continues after
initialization and after any samples have been
discarded.
Manual mode
: The analyzer switches to the
manual mode and waits for user action.
Turbidity compensation
0.10 to 5.00
Factory setting
2.00
Correction factor for turbidity compensation
using the reference wavelength
Sample hose length
Selection
0.10 to 10.00 m
Factory setting
3.00 m
Sample discard mode
Selection
• To process
• To drain
• Do not discard
Factory setting
To process
Specifies whether the content of the sample hose
is discarded after sampling and where.
To process
: The sample is forced back into the
process.
To drain
: The sample is forced into the drain.
Do not discard
: The sample is not discarded and
remains in the sample hose.
Reaction time
60 to 900 s
Factory setting
60 s
Sample retries
Selection
• unlimited
• 1 ... 5
Factory setting
3
10.3.2 Measurement
Menu/Setup/Analyzer/Measurement
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Continuous
Factory setting
Immediate
Immediate
: The analyzer immediately starts
with the measurement cycle once the system
changes to the automatic mode.
Date/time
: The analyzer starts with the
measurement cycle at the set date/time.
Continuous
: The analyzer measures
continuously, without interruption, between the
measurements.
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Measuring interval
0:45 to 24:00 (
hh:mm
)
Factory setting
1:00
Function only available for:
Start condition
=
Immediate
or
Date/time
For configuring the time interval of the
measurement
Summary of Contents for Liquiline System CA80TN