Liquiline System CA80TN
Operation
Hauser
63
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Limits valve hoses
Function
Options
Info
Monitoring
Selection
• Off
• On
Factory setting
On
Diag. code 733
Operating time
Read only
Drain D
1 to 999 d
Factory setting
90 d
Absorption curve recording
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Absorption curve recording
Function
Options
Info
In automatic mode
Selection
• Off
• On
Factory setting
Off
Off
: Recording is only possible in the manual
mode
On
: Recording also takes place in the automatic
mode
Curve
Selection
1 to 7
Factory setting
1
Select the absorption curve to be recorded.
Only one curve can be selected at any one time.
The data are saved in a logbook.